How to Decode CI Audiograms

Decoding CI Audiograms: A Comprehensive Guide to Understanding Cochlear Implant Performance

For individuals with severe to profound hearing loss, cochlear implants (CIs) represent a transformative technology, offering access to sound where traditional hearing aids fall short. However, the journey with a CI extends far beyond the initial surgery and activation. A critical, yet often daunting, aspect of managing CI performance is understanding the audiogram – a graphical representation of hearing ability. For CI users and their families, interpreting these specialized audiograms is paramount to optimizing device settings, tracking progress, and communicating effectively with audiology professionals. This in-depth guide will demystify CI audiograms, providing the knowledge and tools necessary to decode their complex language.

The Foundation: What is a Cochlear Implant Audiogram?

Before diving into the specifics of CI audiograms, it’s essential to grasp the fundamental differences between a conventional audiogram and one designed for cochlear implant users. A standard audiogram measures the softest sounds a person can hear (their hearing thresholds) across a range of frequencies, typically from 250 Hz to 8000 Hz. These thresholds are plotted in decibels (dB HL – Hearing Level), indicating the intensity of sound required for detection.

Cochlear implants, however, bypass the damaged parts of the inner ear and directly stimulate the auditory nerve with electrical signals. Therefore, measuring “hearing thresholds” in the traditional sense becomes irrelevant. Instead, CI audiograms focus on documenting the electrical stimulation levels required for a CI user to perceive sound. These are not measured in dB HL but in Current Units (CU), Clinical Units (CL), or Stimulus Units (SU), depending on the manufacturer and specific programming software. While the exact terminology may vary, the underlying principle remains the same: mapping the electrical dynamic range of the cochlear implant.

The CI audiogram serves as a vital diagnostic and programming tool. It helps audiologists:

  • Establish the electrical dynamic range: This involves identifying the minimum electrical current needed to detect sound (threshold or “T” levels) and the maximum current that can be tolerated comfortably without being too loud (comfort or “C” levels, or Maximum Comfort Levels – MCLs).

  • Verify proper electrode function: Each electrode within the CI array should produce a predictable response. Deviations can indicate issues.

  • Track performance over time: Regular audiograms provide a longitudinal record of how a CI user is adapting to and benefiting from their device.

  • Guide mapping adjustments: The audiogram informs changes to the CI program, ensuring optimal sound perception and comfort.

Deconstructing the CI Audiogram: Key Components and Their Significance

A CI audiogram, while appearing similar to a conventional audiogram with its X and Y axes, carries distinct meanings for each component. Let’s break down its essential elements:

The X-Axis: Electrodes, Not Frequencies

Unlike a traditional audiogram where the X-axis represents frequencies (e.g., 250 Hz, 500 Hz, 1000 Hz), a CI audiogram’s X-axis depicts the individual electrodes within the cochlear implant array. These electrodes are typically numbered, often from the base (highest frequencies) to the apex (lowest frequencies) of the cochlea, though some manufacturers may use a different numbering scheme.

Actionable Insight: The number of electrodes represented on the audiogram corresponds to the number of active electrodes in the CI array, which can vary based on the specific implant model and the individual’s cochlear anatomy. Understanding the electrode numbering convention for your specific implant (e.g., Nucleus, Advanced Bionics, MED-EL) is crucial for accurate interpretation. Often, a higher numbered electrode signifies a more apical (lower frequency) position, and a lower numbered electrode signifies a more basal (higher frequency) position.

The Y-Axis: Electrical Units, Not Decibels

This is perhaps the most significant departure from conventional audiograms. The Y-axis on a CI audiogram represents electrical current levels, measured in various units such as Current Units (CU), Clinical Units (CL), or Stimulus Units (SU). These units are arbitrary scales developed by each manufacturer to quantify the electrical output of the implant.

Actionable Insight: The specific range of electrical units on the Y-axis will depend on the manufacturer. For example, a Nucleus (Cochlear Limited) system might use Current Units ranging from 0 to 255, while an Advanced Bionics system might use Clinical Units from 0 to 240. It’s imperative not to confuse these with decibels. Higher numbers on the Y-axis indicate greater electrical current.

The Plot Points: Threshold (T) and Comfort (C) Levels

The core of the CI audiogram lies in the plotted points, which represent the electrical levels at which the CI user perceives sound. These points are typically categorized into two main types:

  1. Threshold (T) Levels: These represent the minimum electrical current required for the CI user to just barely detect a sound. T-levels are crucial for ensuring audibility of soft sounds. If T-levels are set too high, soft sounds may be inaudible. If set too low, background noise can become overly amplified.

    Actionable Insight: T-levels are often measured using a technique called “auto-T” or by having the patient indicate when they first perceive a sound (a “beep” or “click”). These points are typically represented by a specific symbol (e.g., an ‘X’ or a square) and are usually plotted lower on the Y-axis (meaning less electrical current).

  2. Comfort (C) Levels (or MCLs – Maximum Comfort Levels): These represent the maximum electrical current that the CI user can comfortably tolerate without the sound being unpleasantly loud or painful. C-levels define the upper limit of the electrical dynamic range, ensuring that loud sounds are not overwhelming.

    Actionable Insight: C-levels are typically measured by gradually increasing the electrical current until the patient reports the sound is “loud but comfortable” or “just right.” These points are usually represented by a different symbol (e.g., a circle or a triangle) and are plotted higher on the Y-axis (meaning more electrical current).

The Dynamic Range: The Sweet Spot for Hearing

The vertical distance between the T-level and C-level for each electrode defines the electrical dynamic range (or comfort window) for that specific electrode. This range represents the usable electrical current window within which the CI processes sound.

Actionable Insight: A healthy dynamic range is crucial for optimal speech perception. Too narrow a dynamic range means that there is a limited range of sound intensities that can be comfortably heard, leading to difficulty distinguishing between soft and loud sounds. Too wide a dynamic range can indicate that the T-levels are too high, meaning soft sounds are missed, or the C-levels are too low, meaning loud sounds are not perceived with sufficient intensity. Ideally, the T and C levels should maintain a consistent, appropriate separation across all active electrodes.

Slope and Shape of the “Audiogram”: Interpreting the Pattern

While not a true “slope” in the traditional sense, the pattern of T and C levels across the electrodes provides critical insights.

  • Flat Profile: A relatively flat profile, where T and C levels are consistent across most electrodes, often indicates a well-balanced map and predictable performance.

  • Sloping Profile: If T and C levels gradually decrease or increase across the electrode array, it might reflect the natural tonotopic organization of the cochlea (lower frequencies needing more current due to deeper insertion, or vice-versa depending on the specific anatomy and implant). However, a sharp, uncharacteristic slope or dip could indicate an issue with a particular electrode or a need for mapping adjustments.

  • Notches or Peaks: Sudden drops or spikes in T or C levels at specific electrodes can be significant. A sudden drop in a T-level might mean that electrode is overstimulating, while a sudden rise might mean understimulation. A sudden drop in a C-level could indicate discomfort at lower levels for that specific electrode. These anomalies require careful investigation by the audiologist.

Actionable Example: Imagine a CI audiogram where the C-levels for electrodes 1-5 are around 180 CU, but suddenly drop to 120 CU for electrode 6, then return to 180 CU for electrodes 7-10. This “notch” at electrode 6 suggests that this specific electrode is causing discomfort at much lower current levels than its neighbors. The audiologist might investigate if there’s a localized tissue reaction, an issue with the electrode itself, or if the patient’s perception of loudness is different for that frequency region. They might then lower the C-level for electrode 6 to improve comfort without significantly impacting overall sound quality.

Advanced Considerations in CI Audiogram Interpretation

Beyond the basic components, several advanced factors come into play when meticulously decoding CI audiograms:

Neural Response Telemetry (NRT) or Neural Response Imaging (NRI)

Many modern CI systems incorporate objective measures to assess neural responses to electrical stimulation. These techniques, often called Neural Response Telemetry (NRT) for Cochlear Limited implants, or Neural Response Imaging (NRI) for Advanced Bionics, measure the electrically evoked compound action potential (ECAP) from the auditory nerve.

How it works: A brief electrical pulse is delivered to a specific electrode, and the neural response is recorded. The lowest electrical level at which a measurable neural response is detected is called the NRT/NRI threshold.

Significance: NRT/NRI thresholds provide an objective measure of neural responsiveness, which can be invaluable, especially for young children or individuals who cannot provide reliable behavioral responses. These objective thresholds are often correlated with behavioral T-levels.

Actionable Insight: If NRT/NRI thresholds are significantly different from behavioral T-levels (e.g., NRT threshold is much higher than the behavioral T-level), it might indicate issues with neural integrity, unusual perception, or potentially inaccurate behavioral responses. Audiologists often use NRT/NRI as a guide when setting T-levels, particularly during initial activations or for challenging cases. A consistent gap between NRT/NRI thresholds and behavioral T-levels across electrodes is often expected, but large deviations warrant investigation.

Input Dynamic Range (IDR) and Mapping Strategies

While not directly plotted on the audiogram, the Input Dynamic Range (IDR) and the chosen mapping strategy significantly influence how the CI processes sound and how T and C levels are set.

  • Input Dynamic Range (IDR): This refers to the range of acoustic input levels (in dB SPL) that the CI processor will compress and map into the individual’s electrical dynamic range (between T and C levels). A typical IDR might be 60-70 dB SPL (e.g., from 20 dB SPL to 80 or 90 dB SPL).

  • Mapping Strategy: Different manufacturers and specific programs within a manufacturer offer various sound processing strategies (e.g., ACE, SPEAK, CIS, FSP). These strategies determine how incoming acoustic signals are converted into electrical pulses across the electrode array.

Actionable Insight: The audiologist considers the IDR and mapping strategy when setting T and C levels. For example, if a patient reports that quiet sounds are still too loud even with appropriate T-levels, the audiologist might need to adjust the IDR or fine-tune the gain settings within the mapping strategy. Conversely, if loud sounds are always distorted, the C-levels might need to be adjusted downward or the IDR narrowed. The CI audiogram serves as the visual representation of the output of this complex mapping process.

Electrode Integrity and Impedance

Another crucial aspect, often viewed alongside the audiogram but not directly plotted on it, is electrode impedance. Impedance is a measure of the opposition to electrical current flow in the electrode-tissue interface.

Significance: Regular impedance measurements are vital for monitoring the health of the electrode array.

  • High Impedance: Can indicate a problem with the electrode itself (e.g., a broken wire), a significant amount of fibrous tissue buildup around the electrode, or a bubble of air/fluid inhibiting contact. High impedance can lead to reduced current flow and potentially poor sound quality or dead electrodes.

  • Low Impedance: Extremely low impedance can suggest a short circuit between electrodes or damage to the insulation.

  • Open Circuit: An open circuit indicates a complete break in the electrical pathway, rendering that electrode non-functional.

Actionable Insight: Audiologists typically measure impedance before or during mapping sessions. If an electrode shows significantly high or low impedance, or an open circuit, it will likely be reflected in an abnormal T or C level for that electrode on the audiogram. For instance, an open circuit would result in no response for that electrode, effectively creating a “gap” in the audiogram data or an inability to measure T/C levels. The audiologist might then deactivate that electrode or investigate further.

Practical Scenarios: Decoding Common CI Audiogram Patterns

Let’s explore some common CI audiogram patterns and what they might indicate, along with the potential implications for the CI user.

Scenario 1: Consistent T and C Levels, Good Dynamic Range

Appearance: The T-levels are consistently low across all electrodes, and the C-levels are consistently higher, maintaining a clear, even separation (e.g., T levels around 40-60 CU, C levels around 180-200 CU).

Interpretation: This is generally an ideal audiogram. It suggests:

  • Good audibility: Soft sounds are likely audible.

  • Comfort with loud sounds: Loud sounds are within a comfortable listening range.

  • Well-balanced map: The electrical dynamic range is appropriately set across all electrodes, providing a broad range of sound perception.

  • Stable electrode function: All electrodes appear to be functioning optimally.

Actionable Outcome: The audiologist would likely make minimal adjustments, focusing on fine-tuning for specific listening environments or patient preferences. Continued monitoring would be the primary action.

Scenario 2: High T-Levels Across the Board

Appearance: The T-levels are consistently high across most or all electrodes (e.g., T levels around 100-120 CU), significantly reducing the dynamic range, even if C-levels are normal.

Interpretation: This indicates that the CI user requires a significant amount of electrical current to just barely detect sound. This can lead to:

  • Difficulty hearing soft sounds: Whispers, distant speech, or subtle environmental sounds may be inaudible.

  • Reduced speech understanding: Crucial low-level speech cues might be missed.

  • Fatigue: The brain may be working harder to process sounds at the edge of audibility.

Potential Causes: This could be due to:

  • Under-stimulation: The current map might be providing insufficient electrical input.

  • Neural adaptation: The auditory nerve may require more stimulation over time.

  • Progression of hearing loss: Though less common with CIs, underlying neural changes could contribute.

Actionable Outcome: The audiologist would typically lower the T-levels for the affected electrodes. This aims to restore audibility for softer sounds and expand the usable dynamic range. They would carefully monitor the patient’s comfort and sound quality after the adjustments.

Scenario 3: Low C-Levels, Especially for Loud Sounds

Appearance: The C-levels are consistently low across most or all electrodes (e.g., C levels around 120-140 CU), meaning the user finds sounds uncomfortable at relatively low electrical current.

Interpretation: This indicates that loud sounds are quickly perceived as uncomfortably loud or even painful, leading to:

  • Sound distortion: Loud sounds may sound “clipped” or harsh.

  • Reduced tolerance for loud environments: Noisy places become unbearable.

  • Difficulty enjoying music: Musical dynamics are significantly limited.

Potential Causes:

  • Over-stimulation: The current map might be providing too much electrical input for loud sounds.

  • Neural sensitivity: The individual’s auditory system might be particularly sensitive to electrical stimulation.

  • Recruitment: A phenomenon where the perception of loudness grows very rapidly with small increases in intensity.

Actionable Outcome: The audiologist would primarily lower the C-levels for the affected electrodes. This expands the comfort window by reducing the upper limit of stimulation. The goal is to provide a wider range of comfortable listening for loud sounds without sacrificing audibility of softer sounds.

Scenario 4: A “Hole” or Significant Drop in T/C Levels for Specific Electrodes

Appearance: A sharp, localized drop in both T and C levels for one or a few adjacent electrodes, creating a “notch” in the audiogram. For example, electrodes 5, 6, and 7 show significantly lower T and C levels compared to electrodes 1-4 and 8-10.

Interpretation: This suggests a localized issue or a region of heightened sensitivity.

  • Localized discomfort: The patient might be experiencing pain or discomfort specific to that frequency region, even at lower current levels.

  • Electrode issue: While less common with modern implants, a partially damaged or faulty electrode could manifest this way.

  • Tissue reaction: Scar tissue or inflammation around those specific electrodes could be causing altered perception.

Actionable Outcome: The audiologist would first verify the patient’s perception and comfort for those specific electrodes. They might further lower the T and C levels for the affected electrodes to improve comfort. If the issue persists or worsens, objective measures (NRT/NRI, impedance) would be re-evaluated. In rare cases, an electrode might be temporarily or permanently deactivated if it consistently causes discomfort or malfunction.

Scenario 5: Missing Electrodes or No Response

Appearance: Some electrodes on the audiogram show no plotted T or C levels, or consistently high impedance values are noted.

Interpretation: This indicates that one or more electrodes are not functioning or are not eliciting a reliable response.

  • Open circuit: A complete break in the electrical pathway to that electrode.

  • High impedance: Significant resistance preventing current flow.

  • Non-auditory response: The patient may be perceiving non-auditory sensations (e.g., facial twitching) instead of sound, leading to an inability to set reliable T/C levels.

Actionable Outcome: The audiologist will systematically investigate the cause. Impedance measurements are crucial here. If an electrode is confirmed as non-functional, it will typically be deactivated to prevent potential issues and preserve battery life. The mapping strategy might then be adjusted to compensate for the missing electrode by reassigning frequencies to adjacent, functional electrodes.

The Patient’s Role: Empowering Yourself in the Mapping Process

Understanding CI audiograms isn’t just for audiologists. As a CI user, your active participation and informed feedback are invaluable during mapping sessions.

  1. Be Observant of Your Hearing: Pay close attention to how sounds are perceived in different environments. Are quiet sounds audible? Are loud sounds comfortable? Is speech clear in various settings?

  2. Keep a Journal: Note specific instances of discomfort, distortion, or difficulty hearing certain sounds. Be specific about the type of sound and the environment. “Music sounds buzzy” is more helpful than “my CI isn’t right.”

  3. Communicate Effectively: When describing your perceptions, use clear and descriptive language. Instead of “it’s too loud,” try “it’s uncomfortably sharp” or “it sounds like shouting.”

  4. Ask Questions: Don’t hesitate to ask your audiologist to explain the adjustments they are making and how they relate to your audiogram. Request to see your audiogram and have them explain its current state.

  5. Understand Your Goals: Before each mapping session, have a clear idea of what you want to achieve. Do you want to hear softer sounds, reduce discomfort with loud sounds, or improve speech clarity in noise?

Actionable Example: If you consistently find yourself turning down the volume of your CI processor when someone speaks loudly, this suggests your C-levels might be set too high. During your mapping session, you can explain to your audiologist, “Loud voices, especially in small rooms, feel overwhelming and sometimes cause a ringing sensation.” This specific feedback will guide the audiologist to review and potentially lower your C-levels across the relevant electrodes.

Beyond the Lines: What a CI Audiogram Doesn’t Tell You

While incredibly informative, it’s vital to recognize that a CI audiogram doesn’t provide the complete picture of a CI user’s auditory experience.

  • Speech Understanding in Noise: The audiogram shows the electrical dynamic range, but it doesn’t directly measure how well a person understands speech, especially in challenging listening environments. Speech perception tests (e.g., word recognition in quiet and noise) are crucial for this.

  • Sound Quality and Music Perception: While an optimal audiogram contributes to good sound quality, subjective aspects like music appreciation, timbre perception, and naturalness of sound are complex and not directly represented by T and C levels.

  • Adaptation and Neural Processing: The brain’s ability to adapt to the electrical input and interpret it as meaningful sound is a dynamic process that evolves over time. The audiogram captures a snapshot of the electrical programming but not the brain’s ongoing processing.

  • Patient Satisfaction and Quality of Life: Ultimately, the success of a CI is measured by the patient’s improved quality of life and communication abilities. The audiogram is a tool to optimize the device, but holistic well-being extends beyond the plotted points.

Conclusion

Decoding CI audiograms is an essential skill for cochlear implant users, their families, and audiology professionals. By understanding the distinct axes, the significance of T and C levels, and the patterns they form, individuals can gain a deeper appreciation for their device’s programming and actively participate in their audiological care. This comprehensive guide, free from technical jargon where possible and rich with actionable insights, empowers you to move beyond simply seeing lines on a graph to truly comprehending the electrical language of your cochlear implant. Armed with this knowledge, you can collaborate more effectively with your audiologist, advocating for optimal device settings that enhance your access to sound and improve your communication in the world around you.